Departments of Gastroenterology and Radiology, Cleveland Clinic Foundation, Cleveland, Ohio
Submitted 15 August 2004 ; accepted in final form 21 December 2004
![]() |
ABSTRACT |
---|
![]() ![]() ![]() ![]() ![]() ![]() |
---|
esophageal motility disorders
Impedance monitoring has been recently introduced to measure bolus transit (11) by detecting changes in resistance to current flow when a bolus traverses an electrode pair, and can distinguish liquid (low impedance) from gas (high impedance) boluses. Incorporating electrode pairs on a commercially available solid-state catheter has been accomplished so bolus transit can be assessed simultaneously with manometry.
Videoesophagram recently has shown that bolus entry and exit correlates with impedance fall and rise, respectively (1). In addition, the criteria currently used to define bolus entry and exit were validated, and volume clearance by barium and impedance were almost identical 5 cm above the lower esophageal sphincter (LES) (12). However, normal bolus transit throughout the esophagus has not been validated by the gold standard videoesophagram. Furthermore, changes in impedance with abnormal barium transit, such as stasis and retrograde escape, have not been described. If characteristic impedance changes occur with these abnormal barium transit patterns, this would be useful in interpretation of impedance monitoring. Furthermore, impedance values at baseline, nadir with a bolus, and recovery after bolus exit have not been reported in peer-reviewed literature.
The aim of this work is to correlate bolus transit patterns determined by videoesophagram with bolus transit patterns identified by impedance monitoring and compare them with concomitant manometric findings. In addition, analysis of the barium bolus relative to the radiopaque electrode pair and pressure site will assess the relationship of barium entry and exit to impedance and pressure changes. Finally, baseline impedance values throughout the esophagus will be presented.
![]() |
MATERIALS AND METHODS |
---|
![]() ![]() ![]() ![]() ![]() ![]() |
---|
We studied 15 healthy volunteers (4 men, 11 women; mean age, 43 yr; range, 33.551 yr). They had no esophageal symptoms, and were receiving no medications affecting motility. The protocol was approved by the Cleveland Clinic Foundation Institutional Review Board on October 7, 2003, and all subjects gave written informed consent.
Videoesophagram was performed at a rate of 30 frames/s (OEC General Electric Medical Systems, Salt Lake City, UT). A 9-channel catheter (Fig. 1) allowed simultaneous impedance and manometry monitoring. It was passed per nares such that circumferential solid-state pressure transducers were placed at the LES and 5 cm above the LES, and three unidirectional pressure transducers were 10, 15, and 20 cm above the LES (Konigsberg Instruments, Pasadena, CA). The proximal four pressure sensors also had electrode pairs 2 cm apart embedded around each sensor to record impedance. The catheter was connected to a personal computer with specialized software (Bioview, Sandhill Scientific) for acquisition, analysis, and storage of impedance-manometry data. Synchronization of videoesophagram to the impedance-manometry tracing was achieved by a custom-made device (Sandhill Scientific) that allowed their correlation at a 0.03-s interval.
|
Bolus transit by impedance was first assessed visually for suspected bolus entry and clearance and to determine whether entry was antegrade, simultaneous, or retrograde. Baseline and nadir impedance values were then subjected to computer analysis to confirm entry, and clearing criteria were met. Bolus entry was defined as a fall in impedance 50% from baseline, and bolus exit was returned to that 50% value that defined entry. Normal bolus transit was defined as antegrade bolus entry and clearing at all four sites, as well as maintaining a relatively stable baseline after bolus exit until the next bolus entry.
Esophageal peristalsis was considered normal when propagation was antegrade and velocity <8 cm/s. Esophageal contraction amplitude was normal if >30 mmHg.
Part B. Comparing Barium Transit to Impedance and Esophageal Pressure at a 5-cm Site in the Distal Esophagus
The synchronization device allowed accurate comparison at intervals as short as 0.03 s of barium entry, presence, and exit to the two radiopaque electrodes that are 2 cm apart, as well as the radiopaque pressure site between the two electrodes. To accomplish this, sketchings traced from the fluoroscopy screen were matched to the same interval on the impedance and manometry tracing. We selected a swallow with normal bolus transit by impedance monitoring and videoesophagram, and chose the study segment for analysis to be between the impedance sites 5 and 10 cm above the LES.
Statistics. Descriptive statistics were used for most data in this study. Comparison of impedance baselines was by Mann-Whitney U-test for nonparametric data.
![]() |
RESULTS |
---|
![]() ![]() ![]() ![]() ![]() ![]() |
---|
Correlation between videoesophagram and impedance. We analyzed a total of 86 swallows from 15 healthy volunteers (Table 1). In 97% (83/86) of swallows, the two methods were in agreement with one of three bolus patterns identified in this study: 1) normal bolus transit, 2) bolus stasis, and 3) retrograde escape.
|
|
|
|
Retrograde escape.
Retrograde escape (Fig. 4) was defined by videoesophagram when barium escaped proximally from the bolus into an area of the esophagus previously cleared and by impedance as bolus entry being retrograde rather than antegrade. Retrograde escape by both methods was detected in 17% (14 of 83) of swallows. Despite the retrograde escape, manometry showed normal peristalsis and a peristaltic contraction amplitude 30 mmHg with all swallows.
|
Abnormal transit by only one method. Only 3 of 86 (3%) swallows had divergent results by esophagram and impedance monitoring. One swallow had retrograde escape by barium that was not detected by impedance. Two swallows had stasis (a small residue at site 15 cm above LES) by impedance because recovery after nadir did not return to the 50% value of the original baseline; however, no retention was detected by barium.
Other observations.
In 47 of 86 swallows, impedance detected air in the bolus traversing the esophagus when the following sequence of events occurred (Fig. 5): First, there was an initial, sudden increase in impedance >3,000 (typically to an absolute value >10,000
) due to air at impedance sites as the bolus advanced through the esophagus. Second, in the distal esophagus impedance >7,000
persisted a mean of 3.6 s (range, 2.74.5) as air accumulated and remained there. Third, impedance sharply decreased to < 300
as the barium filled the distal esophagus after air clearing. Liquid total bolus transit time [time (mean ± SE) from liquid entry at the site 20 cm above LES to liquid exit 5 cm above the LES] of boluses with air transit and accumulation was only 0.5 s longer than boluses without air transit (10 ± 0.3 vs. 9.5 ± 0.3; P > 0.05). Videoesophagram confirmed the passage of air throughout the esophagus and accumulation in the distal esophagus. The subjects had no symptoms during the air transit.
|
Part B: Comparing Barium Transit to Impedance and Esophageal Pressure at a 5-cm Site in the Distal Esophagus
The lowest impedance value occurred when the barium bolus covered both electrodes and maximally distended the esophagus (Fig. 6). Impedance began to fall from baseline and began to rise from nadir when the bolus reached the top electrode of the electrode pair on entry and exit, respectively. The 50% impedance value with entry occurred as the barium bolus head approached the bottom electrode having covered the top electrode, and the 50% value with exit occurred just after the bolus tail left the bottom electrode.
|
![]() |
DISCUSSION |
---|
![]() ![]() ![]() ![]() ![]() ![]() |
---|
Normal bolus transit was the most common bolus transit pattern. Not only could no residue of barium be identified throughout the esophagus after barium transit, but impedance baseline after bolus clearance returned to near the prebolus baseline and remained relatively stable until the next bolus entry at all four sites. We found median prebolus baseline 5 cm above the LES to be 2,172 , which is very similar to a preliminary report of 2,227
(10).
A small residue of barium that decreased impedance >50% at the 1520 cm site above the LES occurred with 6 of 86 swallows and a trivial residue without a decrease in impedance in another 9 of 86. Because simultaneous manometry found normal peristalsis in all 15 swallows with small or trivial stasis and normal contraction amplitude with most, we believe this finding is clinically unimportant and should be viewed as a normal variant. However, 5-cm manometric spacing could miss focal peristaltic dysfunction. Our results are similar to Tamhankar et al. (13) who found with videoesophagram that 15% of normals had proximal escape of one-fifth of their swallows. However, this study did not use concomitant manometry. Although Massey et al. (6) and Hewson et al. (3) found proximal escape at the level of the aortic arch when hypotensive peristaltic waves <20 mmHg occurred in a group of patients with esophageal dysmotility, our report shows that proximal stasis can occur despite normal peristalsis and contraction amplitude.
Retrograde escape occurred in 16% of swallows. Videoesophagram found this was because a small portion of the bolus did not clear the distal esophagus before the next swallow. Impedance monitoring found normal bolus exit at all four sites. However, normal bolus exit from the esophagus did not occur because barium residue was in the "gap" below the last impedance site 5 cm above the LES and the LES (see Fig. 3). Impedance monitoring was also abnormal, however, because premature bolus entry was noted at site 4 as the residue escaped retrograde from the most distal esophagus to the site 5 cm above the LES. We found that no swallow had retrograde bolus escape when the interval between swallows was 30 s, and suggest that 30 s is the optimal interval for swallows.
Clinical applications of impedance monitoring have been recently published. A multicenter study using the same catheter as this study recently provided normal values for esophageal function testing. Complete bolus transit of liquid and viscous boluses by impedance was defined as normal bolus entry (50% fall from baseline to nadir) at the site 20 cm above the LES and bolus exit (return to 50% value defining bolus entry) at sites 5, 10, and 15 cm above the LES, and normal was defined as complete bolus transit of 79% liquid swallows and
69% of viscous swallows (15). Our study supports this report by confirming the validity of the 50% of baseline value as defining bolus entry and exit and confirming that the site 20 cm above the LES should not be used to define normal bolus exit as well as providing an explanation (i.e., presence of small bolus residues). However, we did find retrograde escape to 5 cm above the LES resulted in a premature fall in impedance, and feel antegrade entry at sites 5, 10, and 15 cm above the LES should be added to this definition of normal.
The ability of impedance to assess bolus composition allowed us to identify air transit throughout the esophagus, and especially accumulation and persistence for several seconds in the distal esophagus until cleared from the esophagus after LES relaxation. However, because liquid rapidly cleared the esophagus following air emptying, liquid total bolus transit time was minimally delayed. Thus this finding is not clinically important. Other studies have assessed gas transit. Pouderoux et al. (7) used ultrafast computed tomography and demonstrated that swallowing of air (823 ml) occurs along with the liquid bolus, and this mixture disperses in the esophagus with the air ahead of the liquid bolus to the level of the ampulla. Kahrilas et al. (5) identified repeated gas reflux by fluoroscopy that caused chest pain due to failure of upper esophageal sphincter opening and gas venting.
The comparative analysis of individual frames during videoesophagram with impedance and manometry allowed a close examination of the relationship among barium entry, presence, and exit to the radiopaque impedance electrode pair and pressure site. Traditionally, impedance monitoring defines bolus entry as a 50% fall from baseline, and exit as a 50% increase between nadir and the original baseline (15). Our study shows that the 50% values are reasonable choices for bolus entry and exit supporting previous findings of others (12). Furthermore, the 50% value facilitates computer analysis of bolus entry, presence, and exit. The association of barium entry with intrabolus pressure has been previously described, as has the clearance of bolus from the esophagus at the onset of the esophageal contraction (8). This study supports those findings.
Tutuian et al. (14) recently reported combined manometry and impedance monitoring, termed esophageal function test (EFT) in 350 consecutive patients. They found that all scleroderma and achalasia patients and 50% of patients with diffuse esophageal spasm (DES) and ineffective esophageal motility (IEM) had <80% liquid and/or 70% viscous swallows with complete bolus transit as defined previoulsy. Bolus transit patterns were not described for the various motility disorders. However, we have described in a preliminary report using simultaneous barium esophagram, impedance, and manometry that patients with a tight wrap have a characteristic bolus transit pattern by impedance confirmed by videoesophagram. The bolus transits normally to the distal esophagus where most of the barium collects above the wrap; retrograde escape usually
10 cm then occurs before the next swallow due to obstruction. These findings are present despite normal esophageal peristalsis and peristaltic contraction amplitude (4). Thus this is different than retrograde escape in the normal subjects in this study, where escape occurs after the next swallow. We are studying patients with other disorders to determine whether characteristic patterns of abnormal bolus transit may be present. This may give additional information to the traditional interpretation of the EFT that assesses only the number of swallows with incomplete bolus transit.
There are limitations to impedance monitoring. First, the catheter used in this study has a gap between the LES and site 5 cm above the LES, which does not allow detection of a bolus. As a result, only simultaneous videoesophagram could determine that the fall in impedance 5 cm above the esophagus was due to retrograde escape from bolus stasis in the gap rather than gastroesophageal reflux. Additional impedance sites in this critical area would allow this important distinction. Second, impedance baselines were high in these normal volunteers, but may not be in patients with motility disorders. Studies in GERD patients with 24-h impedance pH have shown that retrograde bolus transit (gastroesophageal reflux) determination is difficult and occasionally not possible in the presence of a low baseline with Barrett's esophagus (10). Antegrade bolus transit may also be affected if a low baseline is present, although this has yet to be studied.
In conclusion, videoesophagram and impedance have excellent correlation in identifying bolus transit patterns. Normal transit is the most common transit pattern in healthy subjects. However, retention of barium residue in the proximal esophagus may occur even with normal peristaltic contractions and amplitude, as can retrograde escape from an incompletely cleared bolus if the next swallow is <30 s. These findings support impedance monitoring as a radiation-free transit test, and provide normal bolus transit patterns for comparison to patients with esophageal motility disorders.
![]() |
FOOTNOTES |
---|
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
![]() |
REFERENCES |
---|
![]() ![]() ![]() ![]() ![]() ![]() |
---|
HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
Visit Other APS Journals Online |